Skip to main content

Advertisement

Log in

Stereotactic core needle breast biopsy marker migration: An analysis of factors contributing to immediate marker migration

  • Breast
  • Published:
European Radiology Aims and scope Submit manuscript

Abstract

Objectives

To evaluate breast biopsy marker migration in stereotactic core needle biopsy procedures and identify contributing factors.

Methods

This retrospective study analyzed 268 stereotactic biopsy markers placed in 263 consecutive patients undergoing stereotactic biopsies using 9G vacuum-assisted devices from August 2010-July 2013. Mammograms were reviewed and factors contributing to marker migration were evaluated. Basic descriptive statistics were calculated and comparisons were performed based on radiographically-confirmed marker migration.

Results

Of the 268 placed stereotactic biopsy markers, 35 (13.1%) migrated ≥1 cm from their biopsy cavity. Range: 1–6 cm; mean (± SD): 2.35 ± 1.22 cm. Of the 35 migrated biopsy markers, 9 (25.7%) migrated ≥3.5 cm. Patient age, biopsy pathology, number of cores, and left versus right breast were not associated with migration status (P> 0.10). Global fatty breast density (P= 0.025) and biopsy in the inner region of breast (P = 0.031) were associated with marker migration. Superior biopsy approach (P= 0.025), locally heterogeneous breast density, and t-shaped biopsy markers (P= 0.035) were significant for no marker migration.

Conclusions

Multiple factors were found to influence marker migration. An overall migration rate of 13% supports endeavors of research groups actively developing new biopsy marker designs for improved resistance to migration.

Key Points

• Breast biopsy marker migration is documented in 13% of 268 procedures.

• Marker migration is affected by physical, biological, and pathological factors.

• Breast density, marker shape, needle approach etc. affect migration.

• Study demonstrates marker migration prevalence; marker design improvements are needed.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2

Similar content being viewed by others

Abbreviations

BI-RADS:

Breast Imaging-reporting and data system

SD:

Standard deviation

References

  1. Burbank F, Forcier N (1997) Tissue marking clip for stereotactic breast biopsy: initial placement accuracy, long-term stability, and usefulness as a guide for wire localization. Radiology 205:407–415

    Article  CAS  Google Scholar 

  2. Liberman L, Dershaw DD, Morris EA, Abramson AF, Thornton CM, Rosen PP (1997) Clip placement after stereotactic vacuum-assisted breast biopsy. Radiology 205:417–422

    Article  CAS  Google Scholar 

  3. Reynolds HE (1999) Marker clip placement following directional, vacuum-assisted breast biopsy. Am Surg 65:59–60

    CAS  PubMed  Google Scholar 

  4. Rosen EL, Vo TT (2001) Metallic clip deployment during stereotactic breast biopsy: retrospective analysis. Radiology 218:510–516

    Article  CAS  Google Scholar 

  5. Kruger BM, Burrowes P, MacGregor JH (2002) Accuracy of marker clip placement after mammotome breast biopsy. Can Assoc Radiol J 53:137–140

    PubMed  Google Scholar 

  6. Millman SLMC, Pile-Spellman ER, Higgins HM, Smith SJ (2002) Accuracy of clip placement using the 11 gauge ultrasound-guided hand-held mammotome device (abstr). AJR Am J Roentgenol 178:61

    Article  Google Scholar 

  7. Esserman LE, Cura MA, DaCosta D (2004) Recognizing pitfalls in early and late migration of clip markers after imaging-guided directional vacuum-assisted biopsy. Radiographics 24:147–156

    Article  Google Scholar 

  8. Birdwell RL, Jackman RJ (2003) Clip or marker migration 5–10 weeks after stereotactic 11-gauge vacuum-assisted breast biopsy: report of two cases. Radiology 229:541–544

    Article  Google Scholar 

  9. Burnside ES, Sohlich RE, Sickles EA (2001) Movement of a biopsy-site marker clip after completion of stereotactic directional vacuum-assisted breast biopsy: case report. Radiology 221:504–507

    Article  CAS  Google Scholar 

  10. Parikh J (2004) Ultrasound demonstration of clip migration to Skin within 6 weeks of 11-gauge vacuum-assisted stereotactic breast biopsy. Breast J 10:539–542

    Article  Google Scholar 

  11. Parikh JR (2005) Delayed migration of Gel Mark Ultra clip within 15 days of 11-gauge vacuum-assisted stereotactic breast biopsy. AJR Am J Roentgenol 185:203–206

    Article  Google Scholar 

  12. Philpotts LE, Lee CH (2002) Clip migration after 11-gauge vacuum-assisted stereotactic biopsy: case report. Radiology 222:794–796

    Article  Google Scholar 

  13. Kaplan JA, Grinstaff MW, Bloch BN (2016) Polymer film-nanoparticle composites as new multimodality, non-migrating breast biopsy markers. Eur Radiol 26:866–873

    Article  Google Scholar 

  14. Siegmann KC, Speck S, Baur A et al (2009) Performance of a newly developed clip (Tumark Professional) for MRI-guided lesion localization after MRI-guided vacuum-assisted biopsy--first results. Röfo 181:147–154

    CAS  PubMed  Google Scholar 

  15. Viehweg P, Fabel K, Petzold A, Friedrich K, Laniado M (2007) O-Twist Marker for marking breast cancer under neoadjuvant chemotherapy--first results. Röfo 179:1055–1060

    CAS  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Ashali Jain.

Ethics declarations

Guarantor

The scientific guarantor of this publication is Dr. B. Nicolas Bloch.

Conflict of interest

The authors of this manuscript declare no relationships with any companies, whose products or services may be related to the subject matter of the article.

Funding

The authors state that this work has not received any funding.

Statistics and biometry

No complex statistical methods were necessary for this paper.

Informed consent

Written informed consent was waived by the Institutional Review Board.

Ethical approval

Institutional Review Board approval was obtained.

Study subjects or cohorts overlap

Some study subjects or cohorts have been previously reported at the European Congress of Radiology in 2014 and 2016.

Methodology

• retrospective

• cross-sectional study

• performed at one institution

Additional information

Ashali Jain and Maria Khalid contributed equally to this work.

Rights and permissions

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Jain, A., Khalid, M., Qureshi, M.M. et al. Stereotactic core needle breast biopsy marker migration: An analysis of factors contributing to immediate marker migration. Eur Radiol 27, 4797–4803 (2017). https://doi.org/10.1007/s00330-017-4851-7

Download citation

  • Received:

  • Revised:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00330-017-4851-7